Specifications :
Recording Leads : 12 standard leads : I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6
Recording Channel : 12-ch x 1, 6-ch x 2, 6-ch x 2+1R, 3-ch x 4, 3-ch x 4+1R, 3-ch x 4+3R (R: Rhythm)
Recording Lead Group : Standard / Cabrera / User pre-defined group
Recording Mode : Auto, Manual, Rhythm, Periodic, Auto Start, Elapsed, Calibration(1mV), Copy, TPH test
Full Disclosure Report : Min. 10 sec ~ Max. 10 min. by user set / Manual mode
Recording Length : Min. 3 sec ~ Max. 85 sec. by user select ( Auto mode only )
Recording system : Thermal printer head ( 8 dots/mm) / 203 (vertical) x 508 (horizontal) dpi
Recording Thermal : ECG Paper | 216 (or 210) mm x 30 m, Roll type
Paper : Fax Paper | Grid record
Sensitivity (mm/mV) : 2.5, 5, 10, 20, 40, Auto
Interpretation (option) : Adult / Pediatric, Mean cardiac cycle with segment and measurement
Stress ECG I/F (option) : Wire (RS232/USB) / Bluetooth, Treadmill/Ergometer and PC stress ECG S/W are required
Recording Speed : Auto record | Manual record
(mm/sec) : 5, 10, 12.5, 25, 50, 100 25, 50
Display : Color TFT 7” LCD (Viewing area : 154.08mm x 85.92mm), Dots : 800 x 480
Display Channel (Ch) : 12 (6-ch x 2), 6 (3-ch x 2), 3 (3-ch x 1), 4 (2-ch x 2), 2 (2-ch x 1), 1 (1-ch x 1)
QRS Beat Display & Sound : 30 ~ 240 bpm on real time.
Alarm Display and Sound : Lead-off detection for individual 10 lead, Battery level, Paper empty,HR
Key Operation : Alphanumeric key by membrane and LCD touch screen
Calibration Signal : Sensitivity and frequency response test by using internal 1 mV
Language : English / Indonesia
ECG Data Acquisition : Simultaneous 12-lead ECG
Input Impedance : Greater than 100 MΩ per 10 Hz
A/D Resolution : 16 bits (0.17 μV/bit)
A/D Sampling Rate : 8,000 samples/8-lead/sec
CMRR : Greater than 100 dB
Frequency Response : 0.05 ~ 200 Hz ( -3dB unfiltered )
Time Constant : Greater than 3.2 sec.
Digital Filter : Drift filter | Muscle Filter | AC filter | High frequency filter
0.5 Hz | 25 / 35 Hz | 50 / 60 Hz | 75 / 100 / 150 / 200 Hz
Pace Maker Detection : Option
Arrhythmia Detection : Auto extend recording time to 20 ~ 60 sec by Ventricular Tachycardia and Bradycardia
Data Management : Save / Delete/ Copy / Review / Transfer / Measure
Performance : AAMI standard requirements
Data Transfer I/F : USB memory stick / USB port / Ethernet LAN / Bluetooth (option)/Wifi (option)
Data Save Memory (10 sec) : Internal memory: Standard 4,000 records / Optional SD card (12,000 records per 1GB) /Optional external USB memory stick (12,000 records per 1GB)
BIONICS PCNet S/W : Via Wire USB / Bluetooth / USB memory / LAN/Wifi
HIS Data I/F : HL7 / XML / SCP / PDF /JPG/ PNG / DICOM(PACS)
External Device I/F (option) : Barcode Reader / Magnetic card reader / printer :HP psc1310 HP LeserJet Pro M201 M202 serial
Patient Leakage Current : Less than 10μA
Electrical Safety : Class I (Degree of protection against electrical shock) Defibrillation proof (5000VDC) based
on type CF applied part Complied IEC60601-1 :2005( 3rd edition), IEC60601-2-25:2011, IEC60601-1-2:2007
Operation Power : AC | 100 ~ 240VAC ±10 % ~50/60 Hz free voltage
Battery | Built-in rechargeable battery. NiMH19.2V (2200mA) / 420 reports (150 min)
per 12-ch x 1 (5 sec) auto record with fully charged battery
Dimension (mm) : 333 (L) x 310 (W) x 105 (H)
Weight : 3.3 Kg with battery
Temperature : Operating : +10 ~ +40 ℃ / Transport or Storage : -20 ~ +60 ℃
Relative Humidity : Operating : 30~75 % (non-condensing) / Transport or Storage: 10~90 % (non-condensing)
Atmosphere : Operating : 70kPa ~ 106kPa / Transport or Storage: 50kPa ~ 106kPa